Effect of X-Ray Tube Voltage and Radiation Dose on Modulation Transfer Function in Digital Mammography


Karaaslan M. K., Avc O. M., Muzoǧlu N., GÜNDOĞDU Ö.

2022 Medical Technologies Congress, TIPTEKNO 2022, Antalya, Türkiye, 31 Ekim - 02 Kasım 2022 identifier identifier

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Doi Numarası: 10.1109/tiptekno56568.2022.9960162
  • Basıldığı Şehir: Antalya
  • Basıldığı Ülke: Türkiye
  • Anahtar Kelimeler: digital mammography, modulation transfer function (MTF), radiation dose, X-ray tube voltage, MTF
  • Kocaeli Üniversitesi Adresli: Evet

Özet

© 2022 IEEE.The Modulation Transfer Function (MTF) is a physical metric that explains the relationship between spatial resolution and contrast in the frequency domain for image quality evaluation. This study investigates the effect of the X-ray tube voltage (kVp) and radiation dose on MTF measurement in digital mammography. The MTF was determined for four radiation dose levels at two different kVp values according to the lEC 62220-1-2 standard. In conclusion, there was no significant difference between the results, and the MTF was determined to be independent of kVp and radiation dose.